9 Pass/Fail Test

To Select Source

Press Source to select the channel (channel enabled from CH1 to CH4) to be tested.
During the test, the oscilloscope will judge whether each frame of waveform in the
source complies with the current test mask and those waveforms pass through the
mask area (blue area) is considered as failed.

Mask Range

Users can define their desired test masks.
Press MaskRange  MaskRange to select "Screen Region" or "Cursor Region" for
test. When "Cursor Region" is selected, two gray cursor lines appear on the screen.
At this point, press Cursor A and Cursor B; then, use
to adjust the positions of
the two cursor lines respectively to determine the test range. Or, press CursorAB
and use
to adjust the positions of cursor A and curcor B at the same time. Note
that you can press down
continuously to switch the current cursor.
Press X Mask and Y Mask respectively and turn
, the mask lines appear on the
screen. Press Create Mask to apply the mask currently created. The ranges of the
horizontal and vertical adjustments are from 0.02 div to 4 div and 0.03 div to 4 div
respectively.

Test and Output

Before the test, you can use the following method to set the output mode of the test
results.
Press Msg Display to select "ON" or "OFF". When "ON" is selected, the test results
will be displayed at the upper right corner of the screen.
DS6000 User's Guide
9-3