Waveform Invert; Vertical Scale - Rigol DS6000 Series User Manual

Ds6000 series digital oscilloscope
Hide thumbs Also See for DS6000 Series:
Table of Contents

Advertisement

RIGOL

Waveform Invert

When waveform invert is turned on, the waveform display rotates 180 degree
relative to the ground potential. When the waveform invert is turned off, the
waveform display is normal. Press CH1  Invert to turn waveform invert on or off.

Vertical Scale

The vertical scale adjustment has two modes: "Coarse" and "Fine".
Press CH1  Volts/Div to select the desired mode. Turn the VERTICAL
SCALE to adjust the vertical scale. Turn clockwise to reduce the scale and turn
counterclockwise to increase the scale.
The scale information (e.g.
screen will change accordingly during the adjustment. The adjustable range of the
vertical scale is related to the current probe ratio. By default, the probe attenuation
ratio is 1X and the adjustable range of the vertical scale is from 2 mV/div to 5 V/div.
Coarse adjustment (take counterclockwise as an example): set the vertical scale
in 1-2-5 step namely 2 mV/div, 5 mV/div, 10 mV/div...5 V/div.
Fine adjustment: further adjust the vertical scale within a relatively smaller
range to improve vertical resolution. If the amplitude of the input waveform is a
little bit greater than the full scale and the amplitude would be a little bit lower if
the next scale is used, fine adjustment can be used to improve the amplitude of
waveform display to view the details of the signal.
Note: you can also press VERTICAL
fine adjustments.
2-8
) in the channel label at the bottom of the
SCALE to quickly switch between corse and
2 To Set the Vertical System
DS6000 User's Guide