5 To Trigger the Oscilloscope

Trigger Holdoff

Trigger holdoff could trigger complicated waveforms (e.g. pulse sequence) stably.
Holdoff time is the time the oscilloscope waits to re-enable the trigger circuit. During
the holdoff time, the oscilloscope will not trigger.
Trigger position
Holdoff Time
Press MENU  Setting  Holdoff in the trigger control area (TRIGGER) at the
front panel and use
to change the holdoff time (the default is 100 ns) until the
waveform triggers stably. The adjustable range of the holdoff time is from 100 ns to
10 s.
DS6000 User's Guide
5-6