10 Waveform Recording
3.
Interval
Set the time interval between the frames in waveform recording and the range
available is from 100 ns to 10 s.
4.
Maximum End Frame
The menu shows the maximum number of frames that can be recorded
currently.
As the capacity of the waveform memory is fixed, the more the number of points
of each frame of waveform has, the less the number of waveform frames can be
recorded. Thus, the maximum end frame of waveform recording is decided by
the "Memory Depth". Please refer to the instruction in "Memory Depth" to
select the desired memory depth.
Table 10-1 Memory Depth and Maximum Number of Frames Recorded
Memory Depth
Auto
14k points
140k points
1.4M points
14M points
140M points
DS6000 User's Guide
Maximum End Frame
200064
31986
1999
249
31
2
RIGOL