Function Inspection - Rigol HDO1000 Series Quick Manual

Digital oscilloscope
Hide thumbs Also See for HDO1000 Series:
Table of Contents

Advertisement

To Prepare for Use
5.6

Function Inspection

1. Press the front-panel
is displayed. Click or tap OK to restore the instrument to its factory default
settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown
Figure 5.4
in
3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal
Output Terminal" of the probe, as shown in
4. Set the probe ratio based on the attenuation of the probe, and then click or tap
> Auto.
5. Observe the waveform on the display. In normal condition, you should see a
square waveform similar to the waveform shown in the figure below.
6. Test the other channels in the same way. If you see the waveform, but the square
wave is not shaped correctly as shown above, perform the procedure described in
16
and a prompt message "Restore default settings?"
.
Figure 5.4 Using the Compensation Signal
Figure 5.5 Square Waveform Signal
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.