To Prepare for Use
2. Connect the other terminal of the logic probe to the signal terminal under test.
RIGOL's MSO8000A series has a standard configuration of a logic probe RPL2316.
To apply to different application scenarios, RPL2316 provides three connection
methods to connect the signal under test. For details, refer to
User Guide .
NOTE
Before you connect the logic probe to the signal under test, connect logic probe's
adapter provided in the accessories to its corresponding channel group.
The digital channel input terminal does not support hot plugging. Please do not insert or
pull out the logic probe when the instrument is in power-on state.
7.6

Function Inspection

1. Press
settings?" is displayed. Press OK or tap OK to restore the instrument to its factory
default settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as shown
Figure 7.7
in
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in
14
Figure 7.6 Connect the logic probe
on the front panel, then a prompt message "Restore default
.