CT Analyzer User Manual
6.8.2

Default test card selection

Enabling or disabling test cards for the default Advanced CT test is done using
the Default Test Card Selection option from the Setting Menu. In this case, the
test card selection made in this page becomes effective when starting a new
Advanced CT test.
If you want to enable or disable test cards for your currently active test only,
press the Select Cards soft key in the CT-Object card. In this case, the test card
selection made in this page is immediately effective after pressing the Back soft
key.
To enable or disable a test card, select its entry using the
then press the Add or Remove soft key. Enabled cards are marked with a check
mark, see Figure 6-8.
Figure 6-8: Select Cards page for selecting the default test cards
Some cards require the existence of other cards. This means that if you are adding such a card, the
required other cards are automatically added, too. On the other hand, if you remove a card which is
required by another card, the other card is removed, too. Figure 6-9 shows these dependencies.
CT-Object
Sec. Winding Resistance
requires
Prim. Winding Resistance
requires
requires
Assessment
requires
requires
Figure 6-9: Dependencies of test cards
Press the Back soft key to save your selection and to return to the setting menu. Press the
return to the setting menu without saving the selection.
70
requires
Burden
Comment
cursor keys and
Excitation
requires
Residual Magnetism
requires
How to get there:
Press the Main soft key in
any test card
Main Menu:
- Settings
Select soft key
Setting Menu:
- Default Test Card
Selection
Select soft key
-> Select Cards page
Ratio
ESC
OMICRON
key to