Advertisement

934.xxx
Warning [934.xxx] Reverse power, don't disconnect any wires, don't switch off mains, wait until
power is dissipated. I = xxxxA.
Reason:
Solution:
Safety action: All measurements are stopped until reverse power is dissipated.
935.xxx
Error [935.xxx] No valid CT Analyzer software. Insert a CF-Card with valid Software and press
"Update Firmw.".
Reason:
Solution:
936.xxx
Warning [936.xxx] Corrupt Calibration Data for voltage inputs. Change to tools menu and try to
reset factory calibration. Until update of calibration data the device will be not calibrated.
Reason:
Solution:
937.xxx
Warning [937.xxx] Corrupt Calibration Data for power output. Change to tools menu and try to
update factory calibration. Until update of calibration data the device will be not calibrated.
Reason:
Solution:
938.xxx
Warning [938.xxx] MIF data block 1 corrupt.
Reason:
Solution:
939.xxx
Warning [939.xxx] MIF data block 2 corrupt.
Reason:
Solution:
940.xxx
Warning [940.xxx] CMOS data block corrupt. Check all device settings. Press Clear Values to
work with default values. Press OK to work with corrupt device settings.
Reason:
Solution:
941.xxx
Error [941.xxx] Corrupt factory calibration data for voltage inputs. Contact next Omicron service
center.
Reason:
Solution:
942.xxx
Error [942.xxx] Corrupt factory calibration data for power unit. Contact next Omicron service
center.
Reason:
Solution:
943.xxx
Error [943.xxx] Error overwriting calibration data.
Reason:
Solution:
944.xxx
Warning [944.xxx] Corrupt user text! Change to tools menu and try to update text. Until update of
user text, the device will use default text.
Reason:
Solution:
945.xxx
Error [945.xxx] Corrupt or missing phase calibration data for power unit. Contact the next
OMICRON service center.
Reason:
Solution:
OMICRON
The device receives reverse power of > 20mA.
Wait until the power is dissipated within the device and the error message disappears.
Cannot find valid CT Analyzer software.
Insert a Compact Flash card with valid software (CTAnalyzer.bin) in the directory
A:\Omicron.
The calibration data checksum for the analog inputs is not correct.
Try to update the factory calibration using the tools menu. If this does not solve the
problem, contact your next OMICRON service center.
The calibration data checksum for the power output is not correct.
Try to update the factory calibration using the tools menu. If this does not solve the
problem, contact your next OMICRON service center.
The checksum of the factory settings data is faulty.
Licensing or device settings data are possibly corrupt. Contact your next OMICRON
service center.
The checksum of the factory settings data is faulty.
Licensing or device settings data are possibly corrupt. Contact your next OMICRON
service center.
Device settings data block corrupt.
Check all device settings.
Factory calibration data block for voltage inputs corrupt.
Contact your next OMICRON service center.
Factory calibration data corrupt.
Contact your next OMICRON service center.
Factory calibration data cannot be restored.
Contact your next OMICRON service center.
Checksum for user language support data invalid.
Copy a user text file (CTUser_xxx.bin) to the directory A:\Omicron\ on the Compact Flash
card and try to update the user language support data using the "Update Text" function in
the tools menu.
The phase calibration data are corrupt or missing.
Contact the next OMICRON service center
Error and warning messages