8.2

CT-Object card

The CT-Object card of the Advanced CT Test mode is the most important card and always required for
all types of tests. In this card all necessary settings for a test are done. Some of the settings contained
in the CT-Object card are also shown in other cards for information.
Note: The CT-Object card is the top user interface level. Pressing the
always bring you back to the CT-Object card.
Figure 8-1: CT-Object card with default settings after initializing a new Advanced CT test
8.2.1

Available soft keys

Table 8-2: Soft keys available in the CT-Object card
Soft key
Description
Clears the results of the previous test and enables to start a new test. All measurement
results as well as all the parameters previously determined using the "guesser function"
are cleared. Text entries for "Location" and "Object" remain unchanged.
Saves the test results and test settings to the currently loaded CT Analyzer report file. If
the test results have not been saved yet, the Save As dialog is opened. Saving a file is
described on page 78.
Saves the test results and test settings to a new .xml file. Saving a file is described on
page 78.
Allows to load a test previously performed in the Advanced CT Test mode and stored on
the Compact Flash card in order to check its results, to recalculate the results with
different settings or to use its settings for a new test.
When loading a test stored in the CT Test mode, CT Analyzer will automatically switch to
the CT Test  mode.
Note: Recalculation of the existing CT test results in order to verify the behavior of the CT
at different burdens or primary currents can be performed by changing the burden value
on the CT-Object card or the primary current on the Ratio card. The recalculated results
can then be stored again like for a regular test.
Opens the Select Cards page where you can select the test cards for the current test.
The Select Cards page is described on page 70.
Note: When loading a new CT test from the main menu (using the "New Advanced
CT Test" entry), the test card selection defined in the default test settings is used for this
new test. The card selection performed in the CT-Object card will then be discarded.