Connection For Special Applications; Measurement On A Gapped Core - Omicron CT SB2 User Manual

Ct analyzer
Hide thumbs Also See for CT SB2:
Table of Contents

Advertisement

3.5

Connection for special applications

3.5.1

Measurement on a gapped core

Applicable for CT Test mode and Advanced CT Test mode.
For gapped cores, the position of the primary wire inside the core has a large influence on the ratio
measurement results.
Therefore, in order to obtain correct measurement results, it is very important to arrange the primary wire
during measurement to the same position inside the core as it is during real operation. Depending on the
position of the primary wire inside the core, the measured ratio can differ by up to 20%.
The figure below shows how the ratio error can differ depending on the position of the primary wire inside
the core.
Position 2
Ratio error measured: +5%
Primary wire
Figure 3-5: Ratio error depending on the position of the primary wire inside the gapped core
As shown in Figure 3-5, the measured ratio error may differ considerably depending on the position of
the primary wire. Best measurement results are obtained if the primary wire is positioned exactly to the
center of the core. As an alternative, a copper foil formed to a ring and placed to the inner side of the
core can be used as shown in Figure 3-6.
Figure 3-6: Using a copper foil formed to a ring as primary wire
OMICRON
Position 1
Ratio error measured: +0.5%
Copper foil
Setup and connection
Position 3
Ratio error measured: -15%
Air gap
31