8.2.2

Information fields to be filled by the user

Figure 8-2: Editing the location settings in the CT-Object card
The "Location" and "Object" text fields are only used for reporting and documentation purposes. They
can be filled by the user after the test is finished. These fields provide information about the location of
the CT and the CT itself. The content of these fields is defined in the Location settings page and the
Object settings page, respectively.
To open these settings pages, position the cursor on the "Location" or the "Object" field and then press
the Details soft key or the
Table 8-3: Location and Object settings page
Page
Location settings page
Object settings page
OMICRON
key.
Description
The Location settings page provides the following text fields. Each field
can contain a maximum of 40 numbers or letters.
Company, Country, Station, Feeder:
Information where the CT is installed.
Phase: Phase to which the CT is connected.
IEC-ID: IEC-ID of the CT or freely definable information.
The Object settings page provides the following text fields. Each field
can contain a maximum of 40 numbers or letters.
Manufact.: Manufacturer of the CT under test.
Type: Type number or description of the CT under test.
Serial No.: Serial number of the CT under test.
Core: Number of the tested core.
Tap: Description of the tap (e.g. 1S1-1S3, X1-X4, ...).