Omicron CT SB2 User Manual page 182

Ct analyzer
Hide thumbs Also See for CT SB2:
Table of Contents

Advertisement

CT Analyzer User Manual
501.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
504.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
505.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
510.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
511.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
513.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
514.xxx
No message displayed on the CT Analyzer. Message only displayed on remote PC connected to the
CT Analyzer.
Reason:
Solution:
800.xxx
Warning [800.xxx] Flash card access error. Internal file system error.
Reason:
Solution:
801.xxx
Warning [801.xxx] Flash card access error. Invalid file name or path.
Reason:
Solution:
802.xxx
Warning [802.xxx] Flash card access error. Access denied.
Reason:
Solution:
803.xxx
Warning [803.xxx] Flash card access error. File or folder already exists.
Reason:
Solution:
804.xxx
Warning [804.xxx] Flash card access error. Disk full.
Reason:
Solution:
805.xxx
Warning [805.xxx] Flash card access error. No or corrupt flash card.
Reason:
Solution:
180
The remote interface handler couldn't decipher the command.
Remote interface error. Check the connection between CT Analyzer and PC.
Data transmission error.
Check the connection between CT Analyzer and PC.
Data transmission error.
Check the connection between CT Analyzer and PC.
Remote interface error: Failed to get the required measurement data because the
measurement is still running.
Check the connection between CT Analyzer and PC.
Remote interface error: Failed to unzip the transmitted data block.
Check the connection between CT Analyzer and PC.
The current remote interface command is not allowed.
Make sure that the current CT Analyzer state allows the execution of the command.
Remote interface error: Remote command was submitted with an invalid parameter.
Verify the parameter (index) submitted with the remote interface command.
The file system sent an error message.
Compact Flash card is possibly corrupt. Use a new Compact Flash card.
The file system sent an error message.
The Compact Flash card is possibly corrupt. Try to save the data to a PC and format the
CF card.
The file system sent an error message. You tried to open a "read only" file or a special
directory.
Access the Compact Flash card on a PC and remove the "read-only" file attribute. File
attributes cannot be modified with the CT Analyzer.
The file system sent an error message.
Enter a different file or folder name.
The file system sent an error message.
Delete some files on the Compact Flash card or insert a new CF card.
No or no valid CF card can be found.
Insert a valid Compact Flash card.
OMICRON