Appendix B Module and Battery Replacement Procedures
Installing the RF Module
B.6
Installing the RF Module
The following procedure provides step by step instructions for installing the RF Application
Module on the OneExpert base.
CAUTION
Modules are ESD sensitive and should only be installed, removed and/or serviced by
Qualified Service Personnel.
Mise en Garde
Les modules sont sensibles aux DES et ils doivent seulement être installés, enlevés ou
entretenus par du personnel de service qualifié.
How to Install the CX100 RF Module
1.
Remove the back cover from the Base.
2.
Install the RF Module to the Base by aligning the parts and feeding the backplane
connector on the Base through the connector opening of the RF Module. Verify the
RF Module is flush on the Base before proceeding.
3.
Firmly press the RF Module together with the Base.
4.
Install 6 screws securing the RF Module to the Base
6 in-lbs.
CAUTION
Do not over tighten the screws, doing so may crack the device's cover.
Rev 002
B - 6
Base Cover Removal
Figure B-4 CX100 RF Module Installation Diagram
CX100 ComXpert Operation Manual
Module and Base Alignment
(Figure
B-1). Torque screws to
22144015
August 2023