MALFUNCTION
Display touchscreen is unresponsive.
No receiving a signal at ANT/SWR or DUPLEX connector.
22144015
August 2023
TEST OR INSPECTION
CORRECTIVE ACTION
If a charged, CX100 lithium battery is available, perform the following:
4
Install battery in the test set. Does the test set power on?
Yes, probable cause is failure at AC/DC input connector,
contact TAC.
No, contact TAC.
1
Reboot device.
2
If the device is configure for VNC access, determine if the UI is
operable using a VNC viewing application.
Is test set Receiver Port set to correct input (ANT/SWR or
1
DUPLEX)?
No, set Receiver Port to correct input. If problem persists,
proceed to next step.
Yes, proceed to next step.
CX100 ComXpert Operation Manual
Chapter 8 Care, Maintenance, and Troubleshooting
Troubleshooting
Rev 002
8 - 9