Gather Test Data; Test Uut Power And Frequency; Test Uut Modulation Level And Distortion - Viavi CX100 ComXpert Operation Manual

Handheld radio test set
Hide thumbs Also See for CX100 ComXpert:
Table of Contents

Advertisement

6.3.5 Gather Test Data

6.3.5.1
Viewing Test Data
Open either the Test Setup Summary panel or the Meters panel.
NOTE
If upper and lower limits are being used, recommendation is to view results on the
Meters panel in order to utilize the color coded pass/fail indicators on meters.

6.3.6 Test UUT Power and Frequency

1.
Key the UUT.
2.
Review the RF Power Meter reading along with overall settings.
3.
Review the RF Frequency Error Meter reading.
4.
Unkey the UUT.

6.3.7 Test UUT Modulation Level and Distortion

1.
Open the AF Function Generator panel.
2.
Set AF Gen 1 soft-key to ON.
3.
Open the Test Setup Summary panel.
4.
Key the UUT and hold.
5.
Review the Modulation Meter reading.
6.
Review the Distortion Meter reading along with overall settings
7.
Unkey the UUT.
22144015
August 2023
CX100 ComXpert Operation Manual
Chapter 6 Performing Tests and Measurements
AM Transmitter Testing
Rev 002
6 - 15