Chapter 5 RF Instrument Function Descriptions
RF Test
Soft-key
Pause/Resume
Soft-key
5.2.9.3
RF Analyzer Display Settings
The Display Settings soft-key opens a menu that contains controls and settings that
manage how traces are displayed on plot fields.
Control/Setting
Active Trace
Button
Average (Field)
Rev 002
5 - 34
Table 5-17 RF Analyzer Soft-keys (Continued)
Description
An attention indicator is displayed on the Normalize soft-key
when the system detects residual power on the signal path.
When this condition occurs, Normalize should be performed to
ensure accurate measurements.
When the RF Analyzer is displayed on the UI, the signal trace is
active by default. Pressing the Pause soft-key temporarily
halts data updates to the UI. When this occurs the soft-key
updates to the Resume soft-key.
When the Analyzer is paused, data that is being acquired is
stored in the system's data acquisition buffer, when the
Resume soft-key is pressed, the data being stored in the
buffer is released and sent to the UI.
Table 5-18 RF Analyzer Display Settings
Description
The Active Trace button opens a window that selects the type
of trace(s) to be displayed on the plot field. Multiple traces can
be selected at any given time on each plot.
Max: displays highest signal received
Min: displays lowest signal received
Live: displays live signal trace
Average: displays an averaged signal trace; average is
defined by the Average Field.
The Average field is enabled when Average is selected as the
active trace. The Average field defines the number of traces to
be used to calculate an average signal trace.