MALFUNCTION
No generator output at ANT/SWR or DUPLEX Connector.
No input signal received at Audio In connector.
22144015
August 2023
TEST OR INSPECTION
CORRECTIVE ACTION
Is correct output connector selected on device (ANT/SWR or
1
DUPLEX)?
No, select correct output connector. If problem persists,
proceed to next step.
Yes, proceed to next step.
2
Is RF Generator output turned ON?
No, turn on RF Generator. If problem persists, proceed to
next step.
Yes, proceed to next step.
3
Is RF Generator Frequency set to correct frequency?
No, set RF Generator Frequency to correct frequency. If
problem persists, proceed to next step.
Yes, proceed to next step.
4
Is RF Generator Level set appropriately for UUT?
No, set RF Generator Level to correct setting. If problem
persists, contact TAC.
Yes, contact TAC.
Is input cable properly connected to Audio In connector?
1
No, properly connect the cable. If problem persists, proceed
to next step.
Yes, proceed to next step.
2
Are Audio In settings appropriate for the characteristics of the
incoming audio signal (e.g. Range, Coupling, Input Sensitivity)?
No, adjust settings. If problem persists, contact TAC.
Yes, contact TAC.
CX100 ComXpert Operation Manual
Chapter 8 Care, Maintenance, and Troubleshooting
Troubleshooting
Rev 002
8 - 11