4−12
Loss/Length Method
When Loss/Length Method is selected, the WireScope 350 calculates
the limit based on the length of the cable tested. You can set values for
connectors, loss per connector, splices, and loss per splice. The
connector and splice losses are added to the loss that is calculated by
multiplying the length of the fiber by the loss/Km specification of the
selected cable, which is shown at the bottom of the Setup screen.
When you select Loss/Length Method and press Next on the Loss
Limit Setup screen, the screen changes to the following:
1 To set any of the variable values, press the variable. The numeric
keypad opens.
2 Press the backspace arrow to erase the displayed value, then enter
the new variable value. Press OK. The keypad closes.
3 Press OK. The Fiber Autotest Settings screen displays.