5 On the Fiber Main Menu, press Calibration. The Fiber Calibration
screen displays.
6 Choose Double Ended, and press Next. The Fiber Detection screen
displays.
7 When the message below the diagram reads, "Ready to Calibrate,"
press OK. The Calibration window opens.
8 When the window reads, "Calibration Successful," press OK in the
window. The Fiber Main Menu displays. Remove the couplers and
use the coupler ends of the test jumpers when changing connections.
Leave the test jumper connections to the Fiber SmartProbes intact.
Method B
Method B or the 1-jumper method is usually recommended for
multimode fiber.