Hide thumbs Also See for WireScope 350:
Table of Contents

Advertisement

2 Select either Overall Budget Method or Loss/Length Method,
then press Next. Continue by completing the steps in the matching
section that follows.
Overall Budget Method
Overall Budget Method configures a fixed loss limit value for each
applicable wavelength. If a multimode SmartProbe is installed, you are
asked to specify overall loss budgets for 850 nm and 1300 nm
wavelengths. If a singlemode SmartProbe is installed, you are asked to
specify overall loss budgets for 1310 nm and 1550 nm wavelengths.
When you select Overall Budget Method and press Next on the Loss
Limit Setup screen, the screen changes to the following:
1 To set the loss limit for a listed wavelength, press the wavelength
field. The numeric keypad opens.
2 Press the backspace arrow to erase the displayed value, then enter
the new limit value. Press OK. The keypad closes.
3 Press OK. The Fiber Autotest Settings screen displays.
8r…‡vs’vtÃAvir…Ã8hiyvt
Tr‡‡vtÃVƒÃhÃ6ˆ‡‚‡r†‡
4−11