4. Settings

4-3-4. Load cell operation settings

To open the LOAD CELL screen, touch the buttons in the fol-
lowing order on the Home Screen.
CALIB. & SYSTEM
e
UNIT SHOWN
Select the indicator value unit.
Select the unit that corresponds to the indicator value.
NOTE
The unit shown is next to the indicator value, but it has no
effect on internal calculations.
For example, the calibration value will not change even if
the unit shown is changed from "N" to "kN".
MAX. DISP. VALUE
Set the highest displayed value.
The default value is set at 110% of the rated capacity.
If this value is exceeded, "+FULL" or "–FULL" will appear in a
pop-up message.
ATTENTION
The range that can be set for MAX. DISP. VALUE is 0–32000.
40
SENSOR
LOAD CELL
e
e
Display example
MOVING AVG. NUM.
Set the moving average number for the measured data.
If zero is set, moving average will not be used.
o This will affect the display and judgment of continuous
judgment mode, but the effect will not be applied to
hold and judgment values.
LOW-PASS FILTER
Set the low pass filter cutoff frequency (Hz).
NOTE
When set to "OFF", only the AD converter anti-aliasing
function is enabled, affecting sampling frequencies to the
high limit of the range.
CALIBRATION LOCK
When "ON", calibration values cannot be changed.
o Keep this "ON" normally so that values are not changed
unintentionally.
SENSOR INPUT LOGIC
The sensor input logic can be reversed artificially.
Normally, "STANDARD" should be used.
NOTE
"REVERSED" does not electrically reverse the input.
After changing this setting, always redo zero.
REMOTE SENSING
Whether or not to use remote sense must be set before
connecting a sensor.