1. Names and functions of parts

1-3. Home Screen

When the unit is turned on, the screen last open when it was
turned off reopens (indicator value or waveform screen).

1-3-1. Indicator value display

Indicator value display screen example
a
ZERO (digital zero) button
This forcefully sets the indicator value to 0 (Digital zero
function) (page 15).
b
Setting button ( )
Touch to open the SETTINGS screen (page 27).
c
Measurement result display button
Touch to show measurement results (page 15).
d
Work number
This shows the currently selected work number.
e
Sensor number
This shows the currently selected sensor number.
f
SD card button
Screenshots and measurement results can be saved on
SD cards.
g
Screen switching button
Touch to switch between indicator value and waveform
display.
h
Panel lock indicator
This shows the "Prevent touchscreen operation" status
set with the control signal input terminal (CONTROL
connector).
: Touchscreen locked
: Touchscreen unlocked
12
i
Reset button
This clears hold display.
j
Load unit
The load unit shown is set by the Unit shown item on
the load cell screen (page 40).
k
Judgment statistics
This shows the number of OK and other judgment results.
l
Displacement judgment
This shows the judgment result.
HI
This appears when the hold value exceeds the high
limit value.
OK
This appears when the hold value is no less than the
low limit value and no more than the high limit value.
LO
This appears when the hold value is below the low
limit value.
H/L
This appears when multiple zones are designated and
HI and LO are mixed.
--
This appears when judgment (hold) was not possible.
m
Displacement indicator value
The number of digits shown is the same as the display
value on the displacement sensor calibration screen.
o Touch this during waveform display to open the
DISPLACE. SENSOR screen.
n
Status
CONTINUE: Continuous judgment
WAIT TRG: Waiting to start measurement
REC: Measuring
STOP: Measuring stopped (holding)
o
Load peak and bottom values
This shows the peak and bottom values of load input.
Touch to reset these values.
p
Management values
This shows the high limit, low limit, high high limit and
low low limit setting values for continuous judgment.
o The high high limit and low low limit are shown
when HHLL is enabled (page 45).
NOTE
During measurement (REC), judgment is conducted using
the management values set in the work.