Actual Load Calibration - KYOWA WGA-680A Series Instruction Manual

Instrumentation amplifier
Hide thumbs Also See for WGA-680A Series:
Table of Contents

Advertisement

Table: FUNCTION SELECTING mode and various functions
FUNCTION
SELECTING
MODE
F-01
The "...." is displayed.
Original value
To display the current input value
in mV/V, press the SET key.
F-02
NOTE
Actual load
Before executing the calibration function, make sure to
calibration
execute the DIGITAL ZERO function with no load applied
to the transducer.
<To enable the key code>
<To disable the key code>
Key operation and WGA-680A movement
in SETTING mode
Apply load to the transducer.
Set a desired numeric value
(load value) to be indicated,
and press the SET key.
Range: ±99999 (Initial value: 4000)
Set a decimal point position
and press the SET key.
The "F00000" appears. After enabling or disabling the
key code, the new calibration value appears.
Remove the load.
The "00000" is displayed at the
end.
In this status, set the "00628"
or "11111"and press the SET key.
Then, the calibration value is
updated.
After the settings, the key code is enabled. To update the
calibration value next time, you are required to enter the
key code.
To disable the key code, execute the "F-12".
The "00000" is displayed at the
end.
Press the SET key.
Then, the calibration value is updated.
After the settings, the key code is disabled. You are able to
update the calibration value without entering the key code.
decimal point position
39
IM-A-1048H
Functional description,
supplements
A function that displays the
raw output value of the
transducer.
A function that applies the
actual load to the transducer
and sets the indicated value
at that time.
For details, see "4-4-1.
ACTUAL LOAD
CALIBRATION".
The "00628" and "11111" are
key codes. The key codes
prevents the calibration
values from changing due to
malfunctions.