6-1-4-6. Load number of measurement waveform samples (0100)
If the measurement status of the unit is "Stop", the most recent measurement result will be returned. If the measurement status of the
unit is anything else, an error (NAK) will be returned.
Command
HEX 0×23 0×30 0×30 0×30 0×31 0×30 0×30 0×32 0×31 0×0D
ASCII
#
0
0
Fixed
Response
When the measurement status of the unit is "Stop"
Byte
0
1
2
HEX 0×06 0×30 0×30 0×30 0×31 0×30 0×30
ASCII ACK
0
0
Fixed
When the measurement status of the unit is not "Stop"
Byte
0
1
2
HEX 0×15 0x30 0x31 0x30 0x30 0x0D 0x0A
ASCII NAK
0
1
Status
LOAD, TIME, DISP, BAND
0: Disabled, 1: Enabled
Data
NUM
Number of samples
0
1
0
0
2
Command No.
Checksum
3
4
5
6
7
0
1
0
0
LOAD TIME DISP BAND
Command No.
3
4
5
6
0
0
CR
LF
1
CR
8
9
10
11
12
NUM
Status
6. Communication functions
13
14
15
16
×
×
Data
Checksum
17
18
0×0D 0×0A
CR
LF
85