KYOWA WP-120A Instructions For Use Manual page 39

Instrumentation amplifier
Hide thumbs Also See for WP-120A:
Table of Contents

Advertisement

4-3-3-6. REWRITE TEDS DATA
The current calibration values (rated output and rated capac-
ity) and calibration date will be written to the TEDS memory.
The indicator unit will not be written.
1
Input "9015" and touch the enter button.
NOTE
o This value must be input to prevent accidental data
change.
o Touch the ESC button to cancel and exit setting mode.
2
Input the calibration date, and touch the enter but-
ton. When "Is it okay to execute?" appears, touch
the OK button.
While the calibration value is being written, "Executing"
appears.
When "Executing" disappears, touch the ESC button to exit
setting mode.
TEDS sensors made by our company only support TEDS cali-
bration. They do not support TEDS data writing or TEDS data
restoration.
4-3-3-7. RESTORE TEDS DATA
Use this to restore the factory default calibration value of a sen-
sor to which data was written using the REWRITE TEDS DATA
procedures above.
ATTENTION
If a TEDS sensor has never had its data replaced, trying to
restore TEDS data to it will result in an error because there is no
data to restore.
1
Input "9015".
NOTE
o This value must be input to prevent accidental data
change.
o Touch the ESC button to cancel and exit setting mode.
2
Touch the enter button. "Executing" will appear
while the unit is restoring data from the TEDS
memory.
When loading the TEDS memory restoration data com-
pletes, the rated output (mV/V) and rated capacity are
shown so the values can be checked.
NOTE
Touch the ESC button to cancel and exit setting mode.
3
Touch the EXECUTE button to restore the data from
the TEDS memory. "Executing" appears while this
occurs.
TEDS sensors made by our company only support TEDS cali-
bration. They do not support TEDS data writing or TEDS data
restoration.