6. Communication functions
Data
Load value
10: ± sign, 11–16: Load value (including decimal places)
Byte
10
11
ASCII
±
×
Displacement value
18: ± sign, 19–24: Displacement value (including decimal places)
Byte
18
19
ASCII
±
×
6-1-4-2. Status polling (0006)
This returns the status of the unit.
Command
HEX 0×23 0×30 0×30 0×30 0×30 0×30 0×36 0×32 0×36 0×0D
ASCII
#
0
0
Fixed
Response
Byte
0
1
2
3
HEX 0×06 0×30 0×30 0×30 0×30 0×30 0×36
ASCII ACK
0
0
0
Fixed
Data
CHECK (measurement complete condition)
0: No, 1: Check
o This becomes 1 when measurement completes and becomes 0 after reading once.
ST1 (command status)
0: Ready, 1: Busy, 2: Error, 3: Continuous transmission active, 4: Sending results automatically
ST2 (measurement status)
0: Continue, 1: Wait, 2: Rec, 3: Stop
0000 (4 digits fixed to 0)
SENSOR (current sensor value memory number)
1–4 (1 digit)
WORK (current work number)
1–16 (2 digits)
80
12
13
14
15
16
×
×
×
×
20
21
22
23
24
×
×
×
×
0
0
0
6
2
Command No.
Checksum
4
5
6
7
0
0
6
CHECK ST1
Command No.
×
×
6
CR
8
9
10
11
0×30 0×30 0×30 0×30
ST2
0
0
Data
12
13
14
15
16
0
0
SENSOR
WORK
17
18
19
20
0×0D 0×0A
×
×
CR
LF
Checksum