4. Settings
EXT. & DISP.
In the same manner as with "EXT. & LOAD" above, after an
external signal start condition is realized, measurement will
start when the displacement exceeds the start level (AND
condition).
LOAD 1
Measurement will start when the load increases to above
the start level. Standby will continue after measurement
completes.
LOAD z
Measurement will start when the load decreases to below
the start level. Standby will continue after measurement
completes.
DISPLACE. 1
Measurement will start when the displacement increases
to above the start level. Standby will continue after mea-
surement completes.
DISPLACE. z
Measurement will start when the displacement decreases
to below the start level. Standby will continue after mea-
surement completes.
When in load/displacement 1z standby, STOP will appear with
a yellow background rather than the usual gray to show the
state clearly.
STOP appearance when in standby
54
4-6-1-2. Measurement stopping settings
EXT. SIGNAL
Measurement stops when an external signal becomes on
or when the waveform display STOP button is touched. See
"4-6-4. Zone judgment timing chart" on page 56 for details.
EXT. | LOAD
After an external signal stop condition is realized, measure-
ment will stop when the load stop level meets one of the
following conditions (OR condition).
o If the measurement starting load value ≤ Stop level
Measurement will stop when the load value exceeds
the stop level.
o If the measurement starting load value > Stop level
Measurement will stop when the load value becomes
equal to or less than the stop level.
EXT. | DISP.
In the same manner as with "EXT. | LOAD" above, measure-
ment will stop when an external signal stop condition is
realized or when the displacement crosses the stop level
(OR condition).
EXT. | TIME
Measurement will stop after an external signal stop condi-
tion is realized or the set time has elapsed.
ATTENTION
o When the graph X axis is time:
Even if a stop condition is not realized, measurement will
stop when the time set for "X AXIS FULL SCALE" has elapsed.
o When the graph X axis is displacement:
Even if a stop condition is not realized, measurement will
stop when the displacement set for "X AXIS FULL SCALE" is
realized.
o During measurement, any of the following operations will
stop measurement and cause measured data to be discarded.
• Control signal input terminal number 27 (force reset) is
turned on (CONTROL connector).
• Control signal input terminal number 33 (clear results (reset
measurement results)) is turned on (CONTROL connector).
• The RESET button shown on the screen is touched.
o The graph will begin to be drawn when measurement starts.
o When the X axis is time, data will be saved and the graph
will be drawn with 0 seconds as the measurements start.
o When the X axis is displacement, data will be saved with
0  mm as the measurements start, and the graph will be
drawn according to the indicator values.