Band Judgment - KYOWA WP-120A Instructions For Use Manual

Instrumentation amplifier
Table of Contents

Advertisement

4. Settings

4-5-2. BAND JUDGMENT

Measurement waveforms are compared to comparison wave-
forms to conduct judgment.
Instead of high limit, low limit and other fixed values for contin-
uous judgment, any curved line can be used for comparison.
Band judgment is conducted when "REC" is the status dis-
played on the screen.
High limit
value
¦qStart judgment
4-5-2-1. Band judgment settings
Touch the buttons in the following order on the Home Screen.
WORK
BAND JUDGMENT
e
e
This sets management values for the load and judges whether
these values are exceeded.
To use band judgment, after setting SELECT REF. WAVE to SELECT,
set WAVEFORM COMP. to ENABLE.
o If no reference waveform has been set, the following
"WAVEFORM COMP. " item cannot be set to "ENABLE".
46
Low limit
value
SELECT REF. WAVE
Set the waveform to use for reference.
WAVE SAMPLING
Use measured data for reference.
WAVE LOADING
Load a waveform saved in the unit.