4. Settings
4-6-4-2. Judgment method
Select the hold method used when measurement data is in
the zone.
The following types of holds are possible.
CONSTANT, SAMPLE, PEAK, BOTTOM, P-P, AVERAGE, MAX/MIN,
INFLECTION
NOTE
After selecting MAX/MIN or INFLECTION, touch it again to
open the settings screen.
4-6-4-2-2. Sampling
The measurement value is held when it enters a zone. The hold is released when it leaves the zone.
Load
Judgment output (load)
OFF
HI
ON
OFF
LO
ON
Judgment output (load)
If the hold point meets a HI or LO condition, this becomes ON.
If the hold point is within the management value range, judgment for this zone will be OK.
58
4-6-4-2-1. Constant comparison
When a measurement value is within a zone, it is compared to
the high and low limits and judged.
Judgment output (load)
This becomes ON at the moment that the load meets a HI or
LO condition. Both HI and LO can become ON.
If the load is never outside the zone, the zone judgment will
be OK.
Hold point
Zone interval
HI: high limit value < measurement value
LO: measurement value < low limit value
Time or displacement