KYOWA WP-120A Instructions For Use Manual page 50

Instrumentation amplifier
Table of Contents

Advertisement

4. Settings
Multi-zone judgment (3 points)
This explanation uses settings for conducting peak, bottom
and inflection point holds with multi-zone judgment as an
example. (A displacement sensor is not used in this example.)
Touch the buttons in the following order on the Home Screen
to open the settings screen.
WORK
e
e
Peak hold
Follow the explanation in the previous section to set the peak
hold. In the settings screen above, the following settings have
been changed to explain multi-zone judgment.
f
ZONE RANGE
Select 0.300–0.700 s.
g
LOAD HIGH (HI)/LOW (LO)
Select 40.00–20.00 N.
h
DISPLACEMENT (HI)/(LO)
This is not used.
Bottom hold
b
Zone
Select 2.
c
Zone 2
Select ENABLE.
e
JUDGE. METHOD (judgment method)
Select BOTTOM.
50
ZONE JUDGMENT
f
ZONE RANGE
Select 0.800–1.200 s.
NOTE
Set the range to conduct the hold.
g
LOAD HIGH (HI)/LOW (LO)
Select 25.00–5.00 N.
NOTE
Set the range for OK judgment.
h
DISPLACEMENT (HI)/(LO)
This is not used.
Inflection point hold
b
Zone
Select 3.
c
Zone 3
Select ENABLE.
e
JUDGE. METHOD (judgment method)
Select INFLECTION.
f
ZONE RANGE
Select 1.300–1.800 s.
NOTE
Set the range to conduct the hold.
g
LOAD HIGH (HI)/LOW (LO)
Select 30.00–10.00 N.
NOTE
Set the range for OK judgment.
h
DISPLACEMENT (HI)/(LO)
This is not used.

Advertisement

Table of Contents
loading

Table of Contents