4. Settings
Loading waveforms
Touch the WAVE LOADING button, and touch the OK button.
A list of measurements will be shown. Use the 5 and b
buttons to select the data, and touch the ENT button to load
the waveform.

4-5-3. Zone judgment

For zone judgment, judgment zones for load and time or load
and displacement are set, and hold values are judged to deter-
mine whether or not they are within a zone.
Band judgment is conducted when "REC" is the status dis-
played on the screen.
Zone judgment settings
The range set with range setting is called a zone.
o A maximum of five zones can be set.
o The judgment method is set for each zone.
o Zones can overlap.
48
When the waveform display horizontal axis is time
Set load management values, and conduct judgments.
Load
Load manage-
ment values
(high limit,
low limit)
When the waveform display horizontal axis is displacement
Set load and/or displacement management values, and conduct
judgments.
Load
Load manage-
ment values
(high limit,
low limit)
NOTE
When the waveform display horizontal axis is displacement,
all zone settings will be set to displacement. They cannot be
set to time.
4-5-3-1. Judgment method
o Select the hold method used when measurement data is in
the judgment range.
o If the hold values are within the management value range,
judgment for that zone will be OK.
o If judgment for all zones is OK, the judgment for that mea-
surement will be OK.
o Judgment outputs (displacement) are output only when
the waveform display horizontal axis is displacement.
Depending on the judgment method, however, judgment
output (displacement) might not be output. For details, see
the explanations for each judgment method.
Zone range
Range
Time
Displacement manage-
ment values
(low limit,
high limit)
Zone range
Range
Displacement

Advertisement