6. Communication functions
6-1-4-7. Load measurement waveform (0101)
If the measurement status of the unit is "Stop", the most recent measurement result will be returned. If the measurement status of the
unit is anything else, an error (NAK) will be returned.
Command
HEX 0×23 0×30 0×30 0×30 0×31 0×30 0×31
ASCII
#
0
0
Fixed
Data
TYPE
0: Load, 1: Displacement, 2: Time (in seconds)
NUM
Number of samples
NOTE
The sample numbers start at 0 and increases by one each time sampling occurs.
Data will not be transferred when 0000 is specified.
Responses
When the measurement status of the unit is "Stop"
Initial data
Byte
0
1
2
HEX 0×06 0×30 0×30 0×30 0×31 0×30 0×31
ASCII ACK
0
0
Fixed
Middle data
Byte
+1
+2
+3
+4
HEX
ASCII
±
×
×
Data (no decimal point)
Last data
Byte
+1
+2
+3
+4
HEX
ASCII
±
×
×
Data for sample number NUM-1
(no decimal point)
When the measurement status of the unit is not "Stop"
Byte
0
1
2
HEX 0×15 0x30 0x31 0x30 0x31 0×0D 0×0A
ASCII NAK
0
1
86
0
1
0
1
TYPE
Command No.
3
4
5
6
7
0
1
0
1
±
Command No.
Data for sample number 0 (no decimal point)
+5
+6
+7
×
×
×
×
Transmit results in order until sample number NUM-2
+5
+6
+7
+8
×
×
×
×
×
Checksum
3
4
5
6
0
1
CR
LF
0×2c
,
NUM
Data
8
9
10
11
12
×
×
×
×
×
+9
+10
+11
0×0D 0×0A
×
CR
LF
0×0D
×
×
CR
Checksum
13
×