Waveform Display - KYOWA WP-120A Instructions For Use Manual

Instrumentation amplifier
Table of Contents

Advertisement

q
Load indicator value
This shows the current indicator value. This can also
show the zone judgment hold value after measurement
(page 57).
The number of digits shown is the same as set for the
rated capacity value.
o Touch this during waveform display to open the
LOAD CELL screen.
r
Load judgment
The judgment result is shown in a green rectangle.
HI
This appears when the hold value exceeds the high
limit value.
OK
This appears when the hold value is no less than the
low limit value and no more than the high limit value.
LO
This appears when the hold value is below the low
limit value.
H/L
This appears when multiple zones are designated and
HI and LO are mixed.
--
This appears when judgment (hold) was not possible.

1-3-2. Waveform display

When the X axis is set to time, the Y axis shows load or dis-
placement. The X axis shows time based on the start of
measurement or the change in displacement.
When the X axis is displacement, the Y axis shows load.
Example of screen during band measurement
s
START/STOP button
Touch to manually start and stop measurement.
t
SCALE button
Touch to zoom in/out or move the waveform display
(page 14).
1. Names and functions of parts
u
Waveform
This shows the measurement results as a waveform.
Values between the low limit and the high limit are shown
in white.
Values above the high limit or below the low limit are
shown in red.
v
Starting point
The Y axis is 0 by default, but it will change if the wave-
form is moved up or down (page 14).
The X axis is fixed to 0.
w
Maximum display value
If load and displacement are set for the Y axis, the max-
imum displacement display value will be shown below
the maximum load display value.
Example of screen during multi-zone judgment
x
x
Zone ranges
These are shown by rectangles of different colors.
o The 8 mark shows the hold position for the zone of
the corresponding color.
1-3-2-1. X and Y axis default values
The X axis is set using "X AXIS FULL SCALE" on the SENSOR
screen (page 29).
The Y axis is 110% of the "MAX. DISP. VALUE" on the load cell
screen (page 40).
o The value set with "X AXIS FULL SCALE" is the basis for the
measurement range. Set the full-scale range before setting
band measurement and multi-zone ranges.