KYOWA WP-120A Instructions For Use Manual page 51

Instrumentation amplifier
Table of Contents

Advertisement

For inflection points, further settings must be made to conduct
inflection point holds. After selecting INFLECTION for the judg-
ment method, touch the selection button again to open the
settings screen.
See "4-6-4-2-8-1. Setting items" on page  64 for information
about setting items, and set values suitable for the waveforms
to be measured.
The results measured with the above settings will be as follows.
i
Hold position
i1: Peak hold position
i2: Bottom hold position
i3: Inflection point hold position
j
Zone range display
The range designated with the ZONE RANGE setting
and the LOAD HIGH (HI)/LOW (LO) limits is shown with
a rectangle.
j1: Peak hold zone
j2: Bottom hold zone
j3: Inflection point zone
NOTE
Zone display colors
1: olive, 2: cyan, 3: orange, 4: light yellow, 5: pink
k
Load indicator value
NOTE
o These settings will show the peak value of zone 1.
o To show the hold value of a different zone, set the
indicator value display to that zone.
l
Load judgment
NOTE
For multi-zone judgment cases, if the judgment for all
zones is OK, the judgment will be OK.