6-1-4-8. TEDS command format
Command numbers 6001–6006 are transmitted in this format.
Data is fixed to 8 characters. If the data lacks 8 characters, each opening will be filled with a "0".
Command
Example of TEDS maximum rated capacity (6002)
HEX 0×23 0×30 0×31 0×36 0×30 0×30 0×32 0×32 0×39 0×0D
ASCII
#
0
1
ID number
Response
Example of TEDS maximum rated capacity (6002)
HEX 0×06 0×30 0×31 0×36 0×30 0×30 0×32 0×32 0×30 0×30 0×32 0×30 0×30 0×30 0×30 0×41 0×44 0×0D 0×0A
ASCII ACK
0
1
ID number
Data
The decimal point position is 2, so the rated capacity is 200.00.
6-1-4-9. Maximum rated capacity (6002), maximum rated output (6003)
and sensor impedance (6004) data formats
Byte
+0
+1
+2
+3
HEX
ASCII POS
×
×
Data (no decimal point)
POS (decimal point position)
0: None, 1: 0.0, 2: 0.00 3: 0.000, 4: 0.0000

6-1-5. Timing

s Command cycle w
Setting
PLC
command
WP-120A
sw
tdr: 5000ms maximum command response
twc: 30ms minimum command interval
6
0
0
2
2
Command No.
Checksum
6
0
0
2
2
Command No.
+4
+5
+6
+7
×
×
×
×
×
Execution
command
Reply data
sw
tdr
twc
9
CR
0
0
2
0
0
8 data characters
Reply data
sw
tdr
6. Communication functions
0
0
A
D
CR
Checksum
Polling
Response
LF