4-3-3. TEDS calibration

A TEDS sensor has calibration information, including rated out-
put and rated capacity, stored in its memory.
TEDS calibration reads this calibration information to automati-
cally record calibration values.
TEDS loading (TEDS calibration)
Zero balancing
Unit shown
c
c
4-3-3-1. TEDS
By connecting a sensor that supports IEEE1451.4 Transducer
Electronic Data Sheets (TEDS) to the unit, the rated output
stored in the sensor can be loaded and this can be applied to
calibration of the indicator with this function.
In addition, the value calibrated by the unit itself can be loaded
to the TEDS sensor and its original value can be restored. These
functions are supported by TEDS sensors that comply with the
standard shown in the table below.
Note, however that equipment with TEDS memory includes
both 1kbit and 4kbit devices, but the table below shows only
support when TEDS memory of 4kbit equipment is connected.
TEDS standard
IEEE
1451.4
Bridge
(V0.9)
Sensors
(33)
-
TEDS calibration, TEDS data writing and TEDS data resto-
ration supported
TEDS calibration supported
-
Not supported
Moreover, TEDS sensors made by our company only support
TEDS calibration. They do not support TEDS data writing or
TEDS data restoration.
4. Settings
IEEE1451.4 (V1.0)
Template ID
Strain Gauge
Other
(35)
-