KYOWA WP-120A Instructions For Use Manual page 60

Instrumentation amplifier
Table of Contents

Advertisement

4. Settings
4-6-4-2-5. Peak to peak
With the point when the zone was entered as zero, the maximum difference between the maximum (peak) and maximum nega-
tive-direction (bottom) values for load becomes the hold value when measurement values are in the zone.
Load
Judgment output (load/displacement)
OFF
HI
ON
OFF
LO
ON
Judgment output (load/displacement)
If the hold point meets a HI or LO condition, this becomes ON.
If the hold point is within the management value range, judgment for this zone will be OK.
4-6-4-2-6. Average value
Calculation of the load average begins when measurement values enter the zone.
The average value of the final position in the zone will be held.
Load
Judgment output (load)
OFF
HI
ON
OFF
LO
ON
Judgment output (load)
If the hold point meets a HI or LO condition, this becomes ON.
If the hold point is within the management value range, judgment for this zone will be OK.
60
P
P
P
B
Zone interval
Zone interval
Hold point
Time or displacement
B
P: peak value
B: bottom value
Hold point
Time or displacement
Hold value
Measurement value
Hold value
Measurement value