4. Settings

4-6-4. Zone judgment timing chart

Switch zone
This is enabled when the "ZONE SWITCHING" setting on the
ZONE JUDGMENT screen is set to "EXT. INPUT" (page 57).
The zone section is determined by signal ON and OFF.
HI and LO judgment outputs (load/judgment)
The indicator value during constant comparison or the
hold value at other times is compared to the high and low
limits and judgment is output.
HI: high limit value < load or hold value
LO: load or hold value < low limit value
Both HI and LO can become ON.
OK judgment output timing
This becomes ON when all judgments in the enabled zone
become OK.
Timing example
Load
Switch zone
OFF
ON
Start/stop measurement
Start/stop
OFF
ON
Measurement complete
OFF
ON
Judgment output (load/displacement)
OFF
HI
ON
OFF
OK
ON
OFF
LO
ON
56
4-6-4-1. Setting items
Touch the buttons in the following order on the Home Screen.
Zone 2
Zone 1
WORK
ZONE JUDGMENT
e
e
Zone 3
Time or displacement