NOTE
The above procedures indicate how to adjust the ZERO by using the output signal
(voltage). The WGA-120A is also able to adjust the ZERO by using the output signal
(current). The output signals (current: 4 to 20 mA) move in tandem with the output signals
(voltage: 0 to +10). (For details, see "2. INITIAL SETTINGS".)
4-3 TO ADJUST THE SPAN
4-3-1 By the actual load calibration
Apply the actual (known) load or weight to the transducers. Adjust the value on the digital
voltmeter, connected to the Output terminal (OUT+ and OUT-), by turning the Sensitivity adjuster.