4. Settings

4-5-4. Band + Zone judgment

4-5-4-1. Setting examples
This explanation uses settings for conducting band judgment
and multi-zone judgment (with peak and inflection point
holds) as an example. (A displacement sensor is not used in
this example.)
In the following example, multi-zone judgment settings are
made first, but settings can be made for either band judgment
or multi-zone judgment first.
Touch the buttons in the following order on the Home Screen
to open the settings screen.
WORK
e
e
a
SELECT REF. WAVE
To newly measure a reference waveform, select "WAVE
SAMPLING". After that, save the reference waveform,
referring to "WAVE SAMPLING" on page  47 and "Editing
comparison waveforms" on page 47.
To load a waveform saved in the unit as the reference
waveform, select "WAVE LOADING". After that, save the
reference waveform, referring to "Loading waveforms"
on page  48 and "Editing comparison waveforms" on
page 47.
b
WAVEFORM COMP.
Select ENABLE.
ATTENTION
If no reference waveform has been set, ENABLE cannot
be selected.
52
BAND JUDGMENT
Touch the HOME button to return to the Home Screen, and
touch the buttons in the following order to open the settings
screen.
WORK
ZONE JUDGMENT
e
e
Peak hold
Follow the explanation in "Peak hold" on page  49 to set the
peak hold.
Inflection point hold
Follow the explanation in "Inflection point hold" on page  50
to set the inflection point hold.
NOTE
Work numbers do not have to be set consecutively.
The results measured with the above settings will be as follows.
c
Load judgment
OK will appear when both band judgment and multi-
zone judgment are OK.
NOTE
Judgment output (load OK) of the control signal output
terminal becomes ON when both band judgment and
multi-zone judgment are OK, but band judgment out-
put is the output of only band judgment.