Sensor Settings - KYOWA WP-120A Instructions For Use Manual

Instrumentation amplifier
Table of Contents

Advertisement

Settings in menus that require warnings are shown when, for
example, other settings must also be changed.
Information
Display example
Confirmation
Display example
Users are asked to reconfirm before restoring the unit to
factory default settings and conducting other processes
that cannot be undone.

4-2. SENSOR settings

To open the SENSOR screen, touch the buttons in the follow-
ing order on the Home Screen.
CALIB. & SYSTEM
e
SENSOR MEMORY
This unit can save 4 sets of settings for connected sensors.
SENSOR
e
ATTENTION
Changing the sensor memory could change the decimal
point positions of the parameters saved in the works
depending on the calibration values.
SAMPLING RATE
This sets the sampling frequency.
Y AXIS
This sets the signal shown for the Y axis in waveform display.
Select either "LOAD" or "LOAD & DISPLACE. "
X AXIS
This sets the signal shown for the X axis in waveform
display.
Select either "TIME" or "DISPLACE. "
X AXIS FULL SCALE
This sets the width of the X axis in waveform display.
o 80 ms and 170 ms cannot be selected when the sam-
pling frequency is set to 5 kHz.
D/A OUTPUT SETTING
This sets whether D/A output is voltage output or current
output.
D/A MAX. VOLTAGE
Voltage setting (limiter)
o The output has an extended range of about 10% more
than the set value. The output voltage range is the
same in the negative direction.
For example, when set to 5 V, the D/A output voltage
range will be about −5.5 V to +5.5 V, and +5 V will be
output when the D/A zero and D/A full scale setting
values are added.
D/A ZERO
Set the indicator value to output for D/A Zero (0V voltage
or 4mA current).
D/A FULL SCALE
With the D/A ZERO value as the reference, set the span for
the indicator values output to D/A.
When the value is the D/A ZERO and D/A FULL SCALE set-
ting values added, the voltage of the D/A MAX. VOLTAGE
setting value (20  mA when in electrical current mode) is
output.
LOAD CELL
For information about load cell settings, see "4-3-4. Load
cell operation settings" on page  40 and "4-3. Load cell
calibration" on page 31.
DISPLACE. SENSOR
For information about displacement sensor settings, see
"4-4. Displacement sensor calibration" on page 41.
4. Settings