4-6-2. Continuous judgment timing chart

o Turning the unit on starts continuous judgment.
o Starting measurement stops continuous judgment and
starts judgment as set by the work.
o After measurement completes, touch the Reset button or
set "Clear results (reset measurement results)" to ON for the
control signal input terminal (CONTROL connector) to start
continuous judgment.
Judgment output operation is as follows.
HH: high high limit value < indicator value
HI:
high limit value < indicator value
LO:
indicator value < low limit value
LL:
indicator value < low low limit value
Judgment output example
Load
High high
limit value
High limit value
Low limit value
Low low
limit value
Judgment
output (load)
OFF
HH
ON
OFF
HI
ON
OFF
OK
ON
OFF
LO
ON
OFF
LL
ON
Clear results
Reset
OFF
ON
Start/stop measurement
Start/stop
OFF
ON
ATTENTION
After measurement using another judgment, setting the Clear
results signal to ON will start continuous judgment.
Time or displacement

4-6-3. Band judgment timing chart

To start and stop measurement, use the start/stop button on
the waveform display screen or use "Start/stop measurement"
with the control signal input terminal (CONTROL connector).
The X axis shows time based on the start of measurement or
the change in displacement.
Judgment starts when the measured value exceeds the com-
parison value starting point and is confirmed when it exceeds
the comparison value ending point.
Judgment output example
Load
High limit value
Waveform judgment
output
OFF
HI
ON
OFF
OK
ON
OFF
LO
ON
Start/stop measurement
Start/stop
OFF
ON
Waveform judgment output operation is as follows.
HI: high limit value < load
LO: load < low limit value
Once HI or LO becomes ON, that state is retained and judg-
ment operations stop.
OK becomes ON when the following conditions are fulfilled
after the measurement value exceeds the comparison value
starting point.
Low limit value ≤ load ≤ high limit value
4. Settings
Low limit
Comparison
value
value range
Time or displacement