Procedures Shared By All Calibration Methods - KYOWA WP-120A Instructions For Use Manual

Instrumentation amplifier
Table of Contents

Advertisement

4. Settings
Procedures shared by all calibration
methods
The three calibration methods are equivalent input calibra-
tion, actual load calibration and TEDS calibration. All the
calibration methods have the same procedures before and
after calibration.
An overview of the calibration procedures is shown below.
Unlock calibration values
Remote sense
Calibration
Calibration
w
or actual load calibration
or TEDS calibration
Lock calibration values
Following an explanation of the procedures to be conducted
before and after calibration are explanations of the procedures
for equivalent input calibration, actual load calibration and
TEDS calibration.
ATTENTION
When calibration is conducted, the D/A converter setting will
be initialized automatically according to the calibration value.
32
c
c
Equivalent input calibration
c
Locking and unlocking calibration values
Usually, the unit is used with CALIBRATION LOCK set to ON. This
must be set to OFF before calibration. After calibration, set it
to ON again.
NOTE
To open the LOAD CELL screen, touch the buttons in the fol-
lowing order on the Home Screen.
CALIB. & SYSTEM
e
1
Touch the CALIBRATION LOCK setting value button
on the LOAD CELL screen.
2
Select OFF or ON.
3
Touch the OK button to confirm the setting.
ATTENTION
In order to prevent accidental changes to the calibration value,
set CALIBRATION LOCK to ON after calibration.
NOTE
See "7-2. Setting value list" on page  90 for information about
settings that cannot be changed when CALIBRATION LOCK is
ON.
SENSOR
LOAD CELL