CHAPTER 3 Settings and Measurement
The differences in XY digital display operation are as follows.
Mode
RANGE
SPAN
In RANGE mode, the relationship between the FS range displayed by the AMP setting and the V/DIV range
displayed on the XY digital display are as follows. In addition, the VERNIER setting range in each range is
as shown in the following table.
FS range
20mV
50mV
100mV
200mV
500mV
1V
2V
5V
10V
20V
50V
100V
200V
500V
1000V
Since the RANGE affects the hardware operation (amplifier sensitivity), it cannot be altered during capturing
or replaying.
The RANGE setting is common to the RANGE setting of the Y-T graph.
The display amplitude of the waveform is determined by the range and X-Y dedicated position, and Vernier
settings, without span setting of Y-T graph.
In SPAN mode, the SPAN can be changed within the same range as the Y-T graph.
Set the SPAN in RANGE/SPAN display of XY digital display with ease. In addition, details can adjusted using
the AMP menu setting or DISP menu setting.
It is common to use span setting of Y-T graph.
3-40
RANGE/SPAN
Use the
keys to change
the V/DIV range.
The V/DIV range cannot be
changed during replay.
Use the
keys to change
the Span width.
The Span width cannot be
changed during replay.
Vernier setting
V/DIV range
range
1mV/DIV
100% to 50%
2mV/DIV
100% to 40%
5mV/DIV
100% to 50%
10mV/DIV
100% to 50%
20mV/DIV
100% to 40%
50mV/DIV
100% to 50%
100mV/DIV
100% to 50%
200mV/DIV
100% to 40%
500mV/DIV
100% to 50%
1V/DIV
100% to 50%
2V/DIV
100% to 40%
5V/DIV
100% to 50%
10V/DIV
100% to 50%
20V/DIV
100% to 40%
50V/DIV
100% to 50%
POSITION
Position reference point is 0V
or 0Vrms.
Position reference point is the
center point of the span.
FS range
10mVrms
25mVrms
50mVrms
100mVrms
250mVrms
500mVrms
1Vrms
2.5Vrms
5Vrms
10Vrms
25Vrms
50Vrms
100Vrms
250Vrms
500Vrms
1000Vrms
VERNIER
100 to 50% (40%)
100 to 40%
Vernier setting
V/DIV range
range
0.5mVrms/DIV
100% to 50%
1mVrms/DIV
100% to 40%
2.5mVrms/DIV
100% to 50%
5mVrms/DIV
100% to 50%
10mVrms/DIV
100% to 40%
25mVrms/DIV
100% to 50%
50mVrms/DIV
100% to 50%
100mVrms/DIV
100% to 40%
250mVrms/DIV
100% to 50%
500mVrms/DIV
100% to 50%
1Vrms/DIV
100% to 40%
2.5Vrms/DIV
100% to 50%
5Vrms/DIV
100% to 50%
10Vrms/DIV
100% to 40%
25Vrms/DIV
100% to 50%
50Vrms/DIV
100% to 50%