Pilot Valve Manifold

The Pilot Valve Manifold (AMETEK Part No. 583011901S) is found on the side of
the Power Supply Board (Figures 4-7 and 4-11).

Removing the Pilot Valve Manifold

1. Using a #2 Philips screwdriver, loosen the two (2) captured screws that
2. Disconnect the pilot valve receptacle plugs (Figure 4-11).
3. Use a 1/16-inch slotted screwdriver to remove the two (2) screws at the
4. Detach the tubing from the manifold by holding the tubing retaining col-
Figure 4-11.
3050-AMS Pilot Valve Manifold.
4-18 | 3050-AMS Moisture Analyzer
secure the Pilot Valve Manifold to the sample manifold plate.
side of the manifold (Figure 4-11).
The mounting screws for the two manifolds are different size.
lar and pulling the tubing free. Make sure that you identify the tubing for
proper position when replacing.
Receptacle Plug
Screws
(Use Slotted
Screwdriver)
PN 305804902, Rev A