Chapter 1, Description of Keys

Keys That Begin with the Letter M

Keys That Begin with the Letter M
MEAS RESET
METER
40
MEAS RESET clears the measurement history for all of the instrument's
measurement algorithms (such as the averaging and peak hold functions) to restart
all measurements that are in progress.
Programming MEAS RESET
The HP-IB commands :MEASure:RESet are used to select this function
programmatically.
Example
OUTPUT 714;":MEAS:RES"
This resets all of the active measurements in the Test Set.
The METER function displays measurements graphically. The METER format is
available for most measurements. To determine if the METER format is provided
for a measurement, position the cursor in front of the measurement's units field
and press the knob. If the message "Press ON/OFF, LIMITs, REF, AVG,
METER, or units" is displayed, the METER format is provided.
Programming METER
The HP-IB command :METer is used to select this data function
programmatically.
To Turn the Meter ON and OFF.
Use the :METer:STATe commands to turn the meter ON and OFF. The parameter
can be a 1 or ON to turn the meter on and a 0 or OFF to turn the meter off.
Syntax
:METer:STATe <ON> or <1>
:METer:STATe <OFF> or <0>
Example
OUTPUT 714;"MEAS:RFR:POW:MET ON"
This turns the analog bar-graph meter ON for the TX power measurement.