Chapter 6, Call Processing Subsystem
Using the CALL CONTROL Screen to Test Call Processing Functions
146
S:\HP892XX\USRGUIDE\manual\callscrn.fb

Paging a Mobile Station That Has Not Registered with the Test Set

1. Ensure that the Test Set is in the active state.
Channel" on page 143
2. Connect the mobile station to the Test Set as shown in
Mobile Station to the Test Set," on page
3. Turn on the mobile station.
4. If the CALL CONTROL screen is not displayed, select CALL CNTL under the To
Screen menu. The CALL CONTROL screen will be displayed.
5. Position the cursor on the Display field and select Data.
6. Select the lower MS Id field. Enter the mobile station identification number, either the
phone number or the MIN number depending upon what the upper MS Id field is set to.
7. Select the Page field. The Page annunciator will light while the Test Set pages the
mobile on the forward control channel. If the mobile station responds properly on the
reverse control channel, the message RECC Return will flash momentarily in the
upper right-hand corner of the CALL CONTROL screen. The Access annunciator
will then light while the Test Set sends the mobile station an alert order on the assigned
voice channel. The mobile station should ring.
8. Press the SEND key on the mobile station. The Connect annunciator will light if the
mobile station properly signals the Test Set when the SEND key is pressed. The mobile
station is now connected to the Test Set on the assigned voice channel. The Voice
Channel Assignment fields will be updated: that is, any "-" in the left-hand fields
will be replaced with appropriate information.
See "Turn On The Test Set's Control
for information on how to activate the Test Set.
119.
Figure 17, "Connecting a