Chapter 4, Description of Fields
Fields That Begin with the Letter A
Arm
Atten Hold
168
This field triggers a single measurement. It is only displayed when the
field is set to
Cntl
Single
when the Test Set is in remote operating mode.
HP-IB Examples
"DISP CRXT;TRIG:MODE:RETR SINGle"
sets the remote operation trigger mode to "Single".
"TRIGger:IMMediate"
triggers all active measurements.
Operating Considerations
Sending a command to trigger a measurement will re-start a measurement if there
is one in progress.
When any of the CDMA screens are selected, all measurements are active. This
eliminates the need to display specific screens to get measurement results.
See Also
"Measurement Triggering Process"
HP 8924C User's Guide
Disarm field description, on page 222
Screens Where Field is Present
CDMA CELLULAR MOBILE RECEIVER TEST
CDMA CELLULAR MOBILE TRANSMITTER TEST
Attenuator hold prevents the fixed RF output attenuators from switching in and
out, eliminating the loss of the output signal as the level is changed. This function
is helpful when making squelch measurements.
Operating Considerations
When this function is set to
15 dB around the present
outside the allowed range results in an error message and beep (if the beeper is
on). RF output level accuracy is greatly degraded outside the allowed range.
Screens Where Field is Present
DUPLEX TEST
RF GENERATOR
RX TEST
. This field is not updated to show the Meas Cntl mode
in the Operating Overview chapter of the
, the RF output level is restricted to a range of
On
setting. Attempting to set an amplitude
Amplitude
Meas
S:\hp8924c\REFGUIDE\MANUAL\fields.chp