Chapter 1, Description of Keys
Keys That Begin with the Letter M
NOTE:
When setting the value of the METER Data Function through HP-IB, a
non-Attribute Unit unit-of-measure must be specified in the command string, otherwise the
current Attribute Unit is assumed by the Test Set.
Refer to
Overview chapter of the
NOTE:
When querying the value of the METER data function through HP-IB, the Test Set always
returns numeric values in attribute units, regardless of the current display units or HP-IB units
settings. Numeric values are expressed in scientific notation.
Refer to
chapter of the HP 8924C User's Guide.
42
To Set the Meter High End and Low End Points.
Use the :METer:HEND and :MEter:LEND commands to set the analog bar-graph
meter's high endpoint and low endpoint.
Syntax
:METer:HEND <real number>
:METer:LEND <real number>
Example
OUTPUT 714;"MEAS:RFR:POW:MET:HEND 20"
OUTPUT 714;"MEAS:RFR:POW:MET:LEND 10"
This sets the analog bar-graph meter's high endpoint to 20 watts and the low
endpoint to 10 watts for the TX power measurement.
"To Specify Units-of-Measure for HP-IB Data Transfer"
HP 8924C User's Guide.
To Query the Meter High End and Low End Points.
Use the :METer:HEND? and :MEter:LEND? commands to query the analog bar-
graph meter high endpoint and low endpoint.
Syntax
:METer:HEND?
:METer:LEND?
OUTPUT 714;"MEAS:RFR:POW:MET:HEND?"
ENTER 714;Meter_hi_end
OUTPUT 714;"MEAS:RFR:POW:MET:LEND?"
ENTER 714;Meter_lo_end
This queries the high end point and low end point of the analog bar-graph
meter for the TX power measurement.
"To Specify Units-of-Measure for HP-IB Data Transfer"
the Operating
in
in the Operating Overview
S:\hp8924c\REFGUIDE\MANUAL\keys.chp