HP 8924C Reference Manual page 198

Cdma mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Chapter 4, Description of Fields
Fields That Begin with the Letter C
198
HP-IB Examples
"DISP CTXT;CDMA:TX:POW:MEAS 'Chan Power'"
selects Channel Power measurements.
"DISP CTXT;MEAS:CDM:CHAN?"
queries the channel power measurement.
"DISP CTXT;MEAS:CDM:CHAN:CAL"
calibrates channel power measurements (connect the DUPLEX OUT and RF IN/
OUT ports before sending this command).
Operating Considerations for the Channel Power Measurement
The cross-calibration process is performed by connecting a cable between the
DUPLEX OUT port and the RF IN/OUT port, then selecting the
field. Refer to
Calibrate
Set chapter of the HP 8924C Application Guide. Channel power measurements
must be calibrated whenever any of the following conditions occur:
Uncal annunciator is flashing
5 °C change in operating temperature
Test Set power cycle
Daily
The Average Power measurement, which is used during cross-calibration, is
automatically zeroed during the calibration procedure.
No special measurement triggers are required for Channel Power measurements.
See Also
Avg Power field description, on page 178
Screens Where Field is Present
CDMA CALL CONTROL
CDMA TRANSMITTER POWER RANGE TEST
CDMA CELLULAR MOBILE TRANSMITER TEST
"Calibrating the Test Set"
Power Meas,
in the Calibrating the Test
S:\hp8924c\REFGUIDE\MANUAL\fields.chp

Advertisement

Table of Contents
loading

Table of Contents